Skip to content
ไทย
BWB

Flame Photometer Model: BWB-SUGAR

Product code: MX-21286

In keeping with the Globally accepted published papers from 1977 forward, potassium and sodium detection is now available online 24/7 with a Uniquely configured channel of detection for extreme accuracy along the range of potassium control.

Prices are quoted, not listed. Add this product to your quote request and we reply with pricing and availability, usually within one business day. No payment, no account, no minimum order.

1

Ask about this product on LINE

Prefer to talk? Call +66 99 241 5692, email us or chat on LINE.

Product description

In keeping with the Globally accepted published papers from 1977 forward, potassium and sodium detection is now available online 24/7 with a Uniquely configured channel of detection for extreme accuracy along the range of potassium control.

This winning proprietary format has been improved with a unique adaptation and optimization of the proprietary BWB optical array detection systems, which in this iteration of the BWB Specialty Flame Photometers allows for a single channel of precise and accurate potassium and sodium measurement.

This Specialty Flame Photometer has been requested and designed for the global Sugar Production marketplace.

There were a few traditional Flame Photometer manufacturers over the decades who had served this market and now these robust and fantastic units are coming out of service.

Enter the BWB SUGAR Flame Photometer with specialized potassium and sodium channel detection, supplied with hybridized specific needs multiple point calibration solutions for potassium and sodium. Per our customers’ requests, our Collection Cup provides Real Time sample access for our mixing chamber/detection systems and only requires that the sample be at or near ambient temperature and customer provided ~5ml/minute feed to our unique “Collection Cup” system.

FP/PC APP

This totally unique BWB feature allows multiple installations details to be shared via intranet or internet links to provide “Distributed Control” via a single Control room with staff completely remote from the Refinery site itself. For large Sugar Companies with multiple sites around the world, this allows Expertise from a centralized location anywhere in the world with real time virtually limitless data accumulation and analysis.

.csv files allow users to construct novel and proprietary Excel spread sheets and to share data quickly and efficiently.

Technical Data

Sample rate

3.5 – 7ml/min

Tubing materials

Silicone and Tygon®

Required desk space for Instrument

50cm (H) x 45cm (W) x 45cm (D)*

For Safety reasons the Flame photometer requires 1m of unobstructed space above to allow dissipation of heat from the chimney.

Instrument size

Sugar Flame Photometer

51cm (H) x 38cm (W) x 41cm (D)

(20in x 14in x 16in)

Shipping

62cm (H) x 55cm (W) x 47cm (D)

(24in x 22in x 19in)

Weight

Sugar Flame Photometer – 15.3kg

Shipping – 25kg

Optimal range

Single point calibration

Na 0.1 – 60ppm

K 0.05 – 100ppm

Multi Point Calibration

Na 0.1 – 1000ppm

K 0.05 – 1000ppm

Reproducibility

<1% Coefficient of variability for 20 consecutive samples over ten minutes at concentrations of 100ppm or less. (After instrument stabilization).

Chemical Interference

Na/K/Li = <0.5% to each other when equal in concentration at <100ppm

Limit of detection (LOD) and

limit of quantification (LOQ)

LOD LOQ

Na – 0.03ppm Na – 0.1ppm

K – 0.02ppm K – 0.05ppm

Time to stability

Less than 15 seconds after sample is introduced into the flame

Ions measured

Potassium

Sodium

Interfaces

USB

4-20mA output capable of direct link to SCADA systems

Optional integrated printer

.CSV and .PDF generated reports and files via FP-PC software

Collection Cup providing 24/7 live automated sampling*

Recommended minimum warm up

time based on ambient temperature

21&ordm;C – 40 minutes

Power requirements

100V – 250V AC at 50 or 60Hz

automatically selecting

Fuel requirement

Propane, Butane or Natural Gas*

regulated to 19Bar. Flow rate of O.4l/min

  • with modifications. BWB Technologies recommends either Propane or Butane for optimum results.

Readout

LCD, four line, alpha numeric, back lit.

Specifications

ModelBWB-SUGAR