Hach DR3900 Laboratory VIS Spectrophotometer without RFID, LPV440.99.00002
Hach DR3900 (product number LPV440.99.00002) is a benchtop visible-range spectrophotometer for water analysis in the laboratory, covering 320 to 1100 nm. This product number is the version without RFID technology.
Prices are quoted, not listed. Add this product to your request. We’ll check the specification, pricing and availability, and confirm the delivery details before you order. No payment, no account, no minimum order.
Ask about this product on LINEPrefer to talk? Call +66 99 241 5692, email us or chat on LINE.
Product description
The DR3900 measures transmittance, absorbance and concentration, and has a scanning mode. Hach lists more than 240 preprogrammed methods and room for 100 user programs; when a method is selected, the instrument sets the wavelength automatically.
With Hach TNTplus vial tests, the instrument reads the 2D barcode on the vial to identify the method, then rotates the vial for 10 absorbance readings that are averaged. Hach states that this excludes the effect of scratches and fingerprints on the glass. Hach also states that the 2D barcode supplies the reagent lot number and expiry date, which are logged with each result.
Hach specifies a wavelength accuracy of ±1.5 nm and a photometric measuring range of ±3.0 Abs, both over 340 to 900 nm, with a 5 nm spectral bandwidth and a gas-filled tungsten lamp. Hach lists rectangular cells of 10 mm, 50 mm and 1 inch and round cells of 13 mm, 16 mm and 1 inch as compatible.
The data logger keeps 2000 measured values with result, date, time, sample ID and user ID. Connections are two USB type A ports, one USB type B port and Ethernet, and the display is a 7-inch color touchscreen. Hach lists 25 interface languages, including English, Chinese, Japanese and Korean; Thai is not among them.
RFID features (method updates, sample ID and certificates of analysis read by RFID) are not available on this model. Hach's RFID versions carry other product numbers (LPV440.99.00012 on Hach's US site, LPV440.99.00001 on Hach's UK site), and Hach offers RFID only in certain countries.
Hach supplies the DR3900 with cell adapter A (for 1-inch round and 1 cm square cells), a light shield, a dust cover, a printed basic user manual and a benchtop power supply (110 to 240 V AC, 50/60 Hz) with power cords. Hach's US product page also lists a matched pair of 1-inch square glass sample cells and 115 V and 230 V cords, but the packing list in Hach's 2025 user manuals has no sample cells and names US and GB cords, so confirm the sample cells and plug type in the quotation. Reagents and TNTplus vial tests are not in any of these lists.
For a quotation, give the number of instruments and the parameters you plan to measure, since reagents and vial tests are separate items.
Specifications
| Wavelength Range | 320 to 1100 nm |
|---|---|
| Wavelength Accuracy | ±1.5 nm (wavelength range 340 to 900 nm) |
| Wavelength Reproducibility | ±0.1 nm |
| Wavelength Resolution | 1 nm |
| Wavelength Calibration | Automatic |
| Wavelength Selection | Automatic, based on method selection |
| Spectral Bandwidth | 5 nm |
| Optical System | Reference beam, spectral |
| Light Source | Gas-filled tungsten (visible) |
| Operating Mode | Transmittance (%), absorbance and concentration, scanning |
| Photometric Measuring Range | ±3.0 Abs (wavelength range 340 to 900 nm) |
| Photometric Accuracy | 5 mAbs at 0.0 to 0.5 Abs; 1 % at 0.50 to 2.0 Abs |
| Photometric Linearity | <0.5 % to 2 Abs; ≤1 % at >2 Abs with neutral glass at 546 nm |
| Stray Light | <0.1 % T at 340 nm with NaNO2 |
| Preprogrammed Methods | >240 |
| User Programs | 100 |
| Data Logger | 2000 measured values (result, date, time, sample ID, user ID) |
| Sample Cell Compatibility | Rectangular: 10 mm, 50 mm, 1 inch; round: 13 mm, 16 mm, 1 inch |
| Display | 7-inch (17.8 cm) color TFT touchscreen, WVGA (800 × 480 pixels) |
| Interfaces | USB type A (2), USB type B, Ethernet |
| RFID | Not applicable with this model |
| Power Supply | Benchtop power supply, 110 to 240 V AC, 50/60 Hz |
| Operating Conditions | 10 to 40 °C, max. 80 % relative humidity (non-condensing) |
| Storage Conditions | -40 to 60 °C |
| Enclosure Rating | IP40 |
| Dimensions (H × W × D) | 151 × 350 × 255 mm |
| Weight | 4.2 kg |
| Included | Cell adapter A (1-inch round and 1 cm square cells), light shield, dust cover, printed basic user manual, benchtop power supply with power cords. Hach's US page also lists a matched pair of 1-inch square glass sample cells and 115 V and 230 V cords; Hach's 2025 manuals list no sample cells and give the cords as US and GB types. |
| Warranty | 12 months (Hach) |