Free quotations · Genuine products · Delivery across Thailand and worldwide

Product code: MX-12355
Calcium presence in water supplies results from passage over deposits of limestone, dolomite, gypsum and gypsiferous shale. Its concentration may extend from 0 to several hundred milligrams per liter,depending on its source and treatment.
Prices are quoted, not listed. Add this product to your quote request and we reply with pricing and availability, usually within one business day. No payment, no account, no minimum order.
Prefer to talk? Call +66 99 241 5692, email us or chat on LINE.
The HI758 Checker® HC (Handheld Colorimeter) provides a simple, accurate and cost-effective way to measure calcium. Required for the development of calcium carbonate skeletons of coral reefs, calcium is a crucial component of marine waters. Designed as a more accurate alternative to chemical test kits, the HI758 provides quick, accurate results in a few easy steps. Features at-a-glance Easier to use and more accurate than chemical test kits Dedicated to a single parameter Small size, big convenience Ideal for: Aquaculture, Water Quality, Environmental, Education The HI758 Marine Calcium Checker HC® uses a 100 to 1 dilution for analysis and thus requires deionized or distilled water absent of calcium to function correctly. Calcium can be found in many freshwater sources and this any amount present in the water used at the C1 phase can cause false high readings. In addition, many hobby grade reverse osmosis deionized water systems may not adequately remove calcium. This can vary based on the types of filters used, their effectiveness and frequency of replacement. We recommend using deionized water with low conductivity or pure vapor distilled water commonly found in most pharmacies. Never use tap water to clean your cuvettes. This is especially important for the HI758 Marine Calcium Checker as calcium is commonly present in tap water and residual amounts present in the cuvette will elevate results produced by the Checker. Our HI98703-50 Cuvette Cleaning Solution ensures the glass cuvette used for colorimetric measurements remains clean and free from imperfections. Calcium Checker Troubleshooting Tips Never rinse your cuvettes with tap water or saltwater Use pure vapor distilled water from your local pharmacy at C1 phase. RODI water from home filtration units may not be suitable for this test as calcium can easily permeate through your filter membranes. Clean your cuvettes with distilled or deionized water absent of Calcium that is going to be used for C1 phase dilution. Measure out 9 ml of distilled or deionized water for C1 phase Make sure no excess aquarium sample saltwater is on the tip of the pipette. Only administer the saltwater that is sucked into the pipette tip.
| Range | 200 to 600 ppm |
|---|---|
| Resolution | 1 ppm |
| Accuracy @ 25°C | ±6% of reading |
| Light Source | light emitting diode @ 610 nm |
| Light Detector | Silicon Photocell |
| Environment | 0 to 50 °C (32 to 122 °F); max 95% RH non-condensing |
| Battery Type | (1) 1.5V AAA |
| Auto-off | after 10 minutes of non-use |
| Dimensions | 81.5 x 61 x 37.5 mm (3.2 x 2.4 x 1.5′′) |
| Weight | 64 g (2.25 oz.) |
| Method | adaptation of the zincon method |

Hanna
MX-11176
HI9829-14/15 is for premium quality 10 ppm and 100 ppm (mg/L) Nitrate (as N) calibration standards. It is produced in accordance with ISO 3696/BS3978 standards using high purity salts, deionized water, certified weight-checked balances, and Class A glassware in a temperature controlled environment monitored with…