HI8410 Dissolved Oxygen Controller with 4 to 20mA

Product code: MX-12510

Extended range to 50 mg/L (ppm) Manual single point calibration Selectable 0-20 or 4-20 mA output Low maintenance Galvanic DO probe Backlit LCD Operational mode LED indicators Transparent splash-proof cover included The HI 8410 is a panel mounted dissolved oxygen controller that is used to maintain and monitor the…

Prices are quoted, not listed. Add this product to your request. We’ll check the specification, pricing and availability, and confirm the delivery details before you order. No payment, no account, no minimum order.

Ask about this product on LINE
1

Prefer to talk? Call +66 99 241 5692, email us or chat on LINE.

Product description

Fail Safe Alarm Syste Hanna's exclusive Fail Safe Alarm System protects against problems caused by power supply failure or signal interruption, which are typical risks in industrial environments. An alarm condition is indicated by a red LED, located directly on the front panel of the controller. Automatic Temperature Compensation The Dissolved Oxygen readings are automatically compensated for the temperature effects on the oxygen solubility and membrane permeability. Salinty Compensation The salinity compensation feature allows the determination of Dissolved Oxygen even in salty waters. Simple to Use The instruments are designed with a standard DIN panel mount with membrane keypads and large LCD with backlight on the front, and provide a series of auto-diagnostic functions. Analog Output To enhance troubleshooting and the ability to record data while monitoring, simply attach a recording device to the instrument's 4 to 20 mA output contacts, conveniently located on the front panel, to obtain a copy of the results on demand. Diagnostic Tests The HI 8410 controllers are designed with built-in diagnostic functions to enable the user to check and troubleshoot the instrument. The checks performed are through the front panel keys and can be used to isolate the cause of malfunction.

Specifications

Range0.0 to 50.0 mg/L (ppm) O 2 0 to 600 % O 2 -5.0 to 50.0 °C
Resolution0.1 mg/L or 1% (O 2 ) / 0.1 °C
Accuracy @ 20°C±1% of reading (O 2 ) / ±0.2 °C
Calibrationmanual, one point, in saturated air
Temperature Compensationautomatic, from -5 to 50°C (23 to 122 °F)
Salinity Compensation0 to 51 g/L (resolution 1 g/L)
Probe (not included)HI 76410/4 with 4 m (13.1’) cable or HI 76410/10 with 10 m (32.8’) cable
Recorder Output0 to 20 mA or 4 to 20 mA (isolated)
Setpoint and Alarm Relay1, isolated, 2A, max. 240V, resistive load, 1,000,000 strokes
Setpoint Range1 to 600 % O 2 ; 0.1 to 50.0 mg/L (ppm) O 2
Alarm Range1.0 to 5.0 mg/L (ppm) O 2
Hysteresis Range0.5 to 2.4 mg/L (ppm) O 2
Dosing ControlOFF/AUTO/ON with selection switch
Over Dosing Controladjustable, from 5 min to 60 min with knob or disable by wire strap – on rear panel
Backlightcontinuous on
Power Supply115 VAC ±10% or 230 VAC ±10%; 50/60 Hz
Enclosureflame retardant ABS body and front panel; transparent splash-proof front cover
Environment-10 to 50°C (14 to 122°F); RH max 95% non-conndensing
Panel Cutout141 x 69 mm (5.6 x 2.7’’)
Weight1 kg (2.2 lb.)