1/3
RaddecAsk about this product on LINE
Pyrolyser | Raddec Pyrolyser-6 Trio Gen IV
Product code: MX-21343
Prices are quoted, not listed. Add this product to your quote request and we reply with pricing and availability, usually within one business day. No payment, no account, no minimum order.
1
Prefer to talk? Call +66 99 241 5692, email us or chat on LINE.
Product description
Specifications
| Model | Pyrolyser-6 Trio Gen IV |
|---|---|
| General | Pyrolyser-6 Trio Gen IV |
| Number of fully independent furnace zones | 3 |
| Number of independent sample work-tube | 6 |
| Minimum sample throughput | 6 samples per 8 hour working day for normal heating cycle - faster cycles possible |
| Maximum sample size per tube | Depends on combustibility - up to 10g (dry) fish per tube successfully tested per worktube |
| Typical catalyst lifetime per worktube | 10 g loading lasts about 20 determinations |
| Typical lifetime of silica liners and worktube | ~ 2 years |
| Time for thermally cleaning silica worktube | 3 hours or overnight |
| Overall mass | Approx. 120 kg |
| Instrument dimensions Example bench footprint needed | 1000 x 600 x 800 mm 2300 x 700 x 1000 mm |
| Power demand (other options available on request) | 7 kW 32 A, 220-240 V, 1-phase electrical supply |
| System cooling to aid new cycle of sample loading | Unique fan-assisted cooling for rapid cycling |
| Sample zone temperature control | HMI-PLC system (Siemens and Eurotherm) All three furnace zones independently controllable for maximum flexibility |
| Over-temperature protection | 3 x Eurotherm EPC 2000 |
| Number of user-defined programs | 20 (each with 15 segment programming) |
| Data logger | Yes |
| Gas supplies required | Oxygen and Air at 1 BarG (Nitrogen option also available) |
| Integrated automatic gas switching | Yes - controlled by the PLC system |
| Trapping media for HTO and CO 2 | 1% Nitric acid in water and Carbosorb TM |
| Bubbler trapping efficoencies | >95% for 3 H and 95% 14 C |
| Typical detection limits (2s) - 3 H and 14 C | Nominally 0.010 Bq/g sample (for 5 g sample and a 2 hour count) |