HALO2 Wireless Refillable pH Tester for Lab with Built-in Specialized Electrode

Product code: MX-11639

Universal Connectivity The open-source Bluetooth® protocol allows easy integration with the Hanna Lab App and current Laboratory Information Management System (LIMS). This HALO2 Wireless pH Tester is GLP compliant.

Prices are quoted, not listed. Add this product to your request. We’ll check the specification, pricing and availability, and confirm the delivery details before you order. No payment, no account, no minimum order.

Ask about this product on LINE
1

Prefer to talk? Call +66 99 241 5692, email us or chat on LINE.

Product description

Universal Connectivity The open-source Bluetooth® protocol allows easy integration with the Hanna Lab App and current Laboratory Information Management System (LIMS). This HALO2 Wireless pH Tester is GLP compliant.

IP65 Water Resistant Protection The housing of the pH Tester is rated IP65 meaning it has a high level of protection against particles and dust, and protection against the ingress of water from jets. This makes it ideal for outdoor use.

Pocket-Sized The compact size makes the pH Tester ideal for on-the-go testing. A large LCD display provides at-a-glance readings from any angle and single button operation streamlines the testing process for any user.

General Purpose Glass Formulation General Purpose (GP) glass provides the best response over the entire pH range and can be used for a wide range of applications.

Built-In Temperature Sensor The HALO2 Lab pH Tester (HI9810462) features a built-in temperature sensor for temperature measurement and compensation.

Spheric Glass Tip The spheric bulb is for general purpose use within laboratory environments and applications that measure aqueous solutions due to its large surface area. Other tip shapes include conic for penetration and flat tip for surface measurements.

Cleaning and Disinfection Solution Your HALO2 pH Tester comes with cleaning solution that has been engineered to keep your electrode free from grease and debris. This concentrated combination of detergents cleans your meter quickly and effectively so that your meter stays in optimum condition.

Easy Calibration HALO2 can be placed right into our calibration buffer sachets for easy calibration. Using a Hanna calibration sachet ensures your buffer is always fresh.

Storage Solution Proper storage of your electrode will ensure reliable measurements and a long electrode life. Our storage solution is specially formulated to keep the pH glass hydrated while minimizing organic growth.

Specifications

pH Range0.00 to 14.00 pH
pH Resolution0.01 or 0.1 pH
pH Accuracy±0.02 pH @ 25°C (77°F)
mV RangepH/mV conversion
mV Resolution0.1 or 1 mV
Temperature Range-5.0 to 80.0 °C (23.0 to 176.0 °F)
Temperature Resolution0.1 °C; 0.1 °F
Temperature Accuracy±0.5 °C; ±0.9 °F
CalibrationUp to three points or five points * Automatic buffer recognition with Standard Hanna® buffers (pH 1.68*, 4.01, 7.01, 10.01, 12.45*) or NIST (pH 1.68 *, 4.01, 6.86, 9.18, 12.45*)
Temperature compensationAutomatic (ATC) or Manual (MTC) *
ElectrodeBody material - Glass - High Temp (HT) Junction - Triple ceramic, Reference cell - Double, Ag/AgCl Electrolyte - 3.5M KCl (refillable) Tip/Shape - Spheric, Ø 9 mm (Ø 0.35in.) Outer diameter - 12 mm (0.47in.) Length - 120 mm (4.7in.)
Battery type3V Lithium – CR2032
Battery lifeApproximately 1000 hours (500 hours with Bluetooth enabled)
Environment0 to 50 °C (32 to 122 °F)
CasingIP65 ingress protection
Dimensions51 x 206 x 21 mm (2.0 x 8.1 x 0.8“) / 60 g (2.1 oz.)
Weight60 g (2.1 oz.)
Supplied withpH 4.01 buffer solution, 20 mL sachet (2 pcs.), pH 7.01 buffer solution, 20 mL sachet (2 pcs.), Electrode cleaning solution, 20 mL sachet (2 pcs.), Electrode storage solution, 13 mL dropper bottle (1 pc.), Electrolyte refill solution, 30 mL, Pipette, 3V Lithium battery – CR2032, Battery safety insert, Quick reference guide with instrument quality certificate
Notes* Available with Hanna Lab App ** Measuring outside the recommended operating temperature range may damage the gel electrolyte and void product warranty. Note: The tester can display measurements from –2.00 to 16.00 pH. Measurements outside of the