From aluminum to zinc, the HI83300 benchtop photometer measures 37 different key water quality parameters using 60 different methods. This photometer features an innovative optical system that uses LEDs, narrow band interference filters, focusing lens and both a silicon photodetector for absorbance measurement and a reference detector to maintain a consistent light source ensures accurate and repeatable photometric readings every time.A digital pH electrode input allows the user to measure pH by a traditional glass electrode. The digital pH electrode has a built in microchip within the probe that stores all of the calibration information. Having the calibration information stored in the probe allows for hot swapping of pH electrodes without having to recalibrate. All pH measurements are automatically compensated for temperature variations with a built in thermistor located in the tip of the sensing bulb for fast and accurate temperature measurement.The HI83300 offers an absorbance measuring mode that allows for CAL CheckTM standards to be used to validate the performance of the system. The absorbance mode allows the user to select one of the five wavelengths of light (420 nm, 466 nm, 525 nm, 575 nm, and 610 nm) to measure and plot their own concentration versus absorbance mode. This is useful for users with their own chemical method and for educators to teach the concept of absorbance by using the Beer-Lambert Law.Two USB ports are provided for transferring data to a flash drive or computer and to use as a power source for the meter. For added convenience and portability the meter can also operate on an internal 3.7 VDC Lithium-polymer rechargeable battery.Please note, reagents are not included and can be purchased here.

Specifications

pH RangePhotometer: 6.5 to 8.5 pH pH electrode: -2.00 to 16.00 pH
pH ResolutionPhotometer: 0.1 pH pH electrode: 0.1 pH
pH AccuracyPhotometer: ±0.1 pH pH electrode: ±0.01 pH
pH CalibrationAutomatic one or two point calibration with one set of standard buffers available (4.01, 6.86, 7.01, 9.18, 10.01)
pH Temperature CompensationAutomatic (-5.0 to 100.0 o C; 23.0 to 212.0 o F); limits reduced based on the pH electrode used
pH CAL Check (electrode diagnostics)clean electrode and check buffer/check probe displayed during calibration
pH MethodPhotometer: phenol red
pH-mV Range±1000 mV
pH-mV Resolution0.1 mV
pH-mV Accuracy±0.2 mV
Oxygen, Dissolved Range0.0 to 10.0 mg/L (as O 2 )
Oxygen, Dissolved Resolution0.1 mg/L
Oxygen, Dissolved Accuracy±0.4 mg/L ±3% of reading
Oxygen, Dissolved MethodAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Azide modified Winkler method
Absorbance Range0.000 to 4.000 Abs
Absorbance Resolution0.001 Abs
Absorbance Accuracy+/-0.003Abs @ 1.000 Abs
Alkalinity RangeFreshwater: 0 to 500 mg/L (as CaCO 3 ) Seawater: 0 to 500 mg/L (as CaCO 3 )
Alkalinity Resolution1 mg/L
Alkalinity Accuracy±5 mg/L ±5% of reading
Alkalinity MethodColorimetric method
Aluminum Range0.00 to 1.00 mg/L (as Al 3+ )
Aluminum Resolution0.01 mg/L
Aluminum Accuracy±0.04 mg/L ±4% of reading
Aluminum MethodAdaptation of the aluminon method
Ammonia RangeLow Range: 0.00 to 3.00 mg/L (as NH 3 -N) Medium Range: 0.00 to 10.00 mg/L (as NH 3 -N) High Range: 0.0 to 100.0 mg/L (as NH 3 -N)
Ammonia Resolution0.01 mg/L; 0.1 mg/L
Ammonia AccuracyLow Range: ±0.04 mg/L ±4% of reading Medium Range: ±0.05 mg/L ±5% of reading High range: ±0.5 mg/L ±5% of reading at 25 °C
Ammonia MethodAdaptation of the ASTM Manual of Water and Environmental Technology, D1426-92, Nessler method
Anionic Surfactants Range0.00 to 3.50 mg/L (as SDBS)
Anionic Surfactants Resolution0.01 mg/L
Anionic Surfactants Accuracy±0.04 mg/L ±3% of reading
Anionic Surfactants MethodAdaptation of the USEPA method 425.1 and Standard Methods for the Examination of Water and Wastewater, 20th edition, 5540C, Anionic Surfactants as MBAS.
Bromine Range0.00 to 8.00 mg/L (as Br 2 )
Bromine Resolution0.01 mg/L
Bromine Accuracy±0.08 mg/L ±3% of reading
Bromine MethodAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, DPD method.
Calcium RangeFreshwater: 0 to 400 mg/L (as Ca 2+ ) Seawater: 200 to 600 mg/L (as Ca 2+ )
Calcium Resolution1 mg/L
Calcium AccuracyFreshwater: ±10 mg/L ±5% of reading Seawater: ±6% of reading
Calcium MethodFreshwater: Adaptation of the Oxalate method Seawater: Adaptation of the Zincon method
Chlorine Dioxide Range0.00 to 2.00 mg/L (as ClO 2 )
Chlorine Dioxide Resolution0.01 mg/L
Chlorine Dioxide Accuracy±0.10 mg/L ±5% of reading
Chlorine Dioxide MethodAdaptation of the Chlorophenol Red method.
Total Chlorine Range0.00 to 5.00 mg/L (as Cl 2 ) Ultra Low Range: 0.000 to 0.500 mg/L (as Cl 2 ) Ultra High Range: 0 to 500 mg/L (as Cl 2 )
Total Chlorine Resolution0.01 mg/L; 0.001 mg/L; 1 mg/L
Total Chlorine Accuracy±0.03 mg/L ±3% of reading Ultra Low Range: ±0.020 mg/L ±3% of reading Ultra High Range: ±3 mg/L ±3% of reading
Chlorine MethodAdaptation of the EPA 330.5 DPD method Free Chlorine (ULR) & Total Chlorine (UHR): Adaptation of the Standard Methods for Examination of Water and Wastewater, 20th edition, 4500-Cl
Chromium, Hexavalent RangeLow Range: 0 to 300 μg/L (as Cr 6+ ) High Range: 0 to 1000 μg/L (as Cr 6+ )
Chromium, Hexavalent Resolution1 μg/L
Chromium, Hexavalent AccuracyLow Range: ±1 μg/L ±4% of reading High Range: ±5 μg/L ±4% of reading
Chromium, Hexavalent MethodAdaptation of the ASTM Manual of Water and Environmental Technology, D1687-92, Diphenylcarbohydrazide method.
Color, Water Range0 to 500 PCU (Platinum Cobalt Units)
Color, Water Resolution1 PCU
Color, Water Accuracy±10 PCU ±5% of reading
Color, Water MethodAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Colorimetric Platinum Cobalt method.
Copper RangeLow Range: 0.000 to 1.500 mg/L (as Cu 2+ ) High range: 0.00 to 5.00 mg/L (as Cu 2+ )
Copper Resolution0.001 mg/L; 0.01 mg/L
Copper AccuracyLow Range: ±0.01 mg/L ±5% of reading High Range ±0.02 mg/L ±4% of reading
Copper MethodAdaptation of the EPA bicinchoninate method
Cyanuric Acid Range0 to 80 mg/L (as CYA)
Cyanuric Acid Resolution1 mg/L
Cyanuric Acid Accuracy±1 mg/L ±15% of reading
Cyanuric Acid MethodAdaptation of the turbidimetric method
Fluoride RangeLow Range: 0.00 to 2.00 mg/L (as F – ) High range: 0.0 to 20.0 mg/L (as F – )
Fluoride Resolution0.01 mg/L; 0.1 mg/L
Fluoride AccuracyLow Range: ±0.03 mg/L ±3% of reading High Range: ±0.5 mg/L ±3% of reading
Fluoride MethodAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, SPADNS method
Hardness, Total RangeLow Range: 0 to 250 mg/L (as CaCO 3 ) Medium Range: 200 to 500 mg/L (as CaCO 3 ) High Range: 400 to 750 mg/L (as CaCO 3 )
Hardness, Total Resolution1 mg/L
Hardness, Total AccuracyLow Range: ±5 mg/L ±4% of reading Medium Range: ±7 mg/L ±3% of reading High Range: ±10 mg/L ±2% of reading
Hardness, Total MethodAdaptation of the EPA recommended method 130.1
Hardness, Calcium Range0.00 to 2.70 mg/L (as CaCO 3 )
Hardness, Calcium Resolution0.01 mg/L
Hardness, Calcium Accuracy±0.11 mg/L ±5% of reading
Hardness, Calcium MethodAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Calmagite method
Hardness, Magnesium Range0.00 to 2.00 mg/L (as CaCO 3 )
Hardness, Magnesium Resolution0.01 mg/L
Hardness, Magnesium Accuracy±0.11 mg/L ±5% of reading
Hardness, Magnesium MethodAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, EDTA Colorimetric method
Hydrazine Range0 to 400 μg/L (as N 2 H 4 )
Hydrazine Resolution1 μg/L
Hydrazine Accuracy±4% of full scale reading
Hydrazine MethodAdaptation of the ASTM Manual of Water and Environmental Technology, method D1385-88, p-Dimethylaminobenzaldehyde method
Iodine Range0.0 to 12.5 mg/L (as I 2 )
Iodine Resolution0.1 mg/L
Iodine Accuracy±0.1 mg/L ±5% of reading
Iodine MethodAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, DPD method
Iron RangeLow Range: 0.000 to 1.600 mg/L (as Fe) High Range: 0.00 to 5.00 mg/L (as Fe)
Iron Resolution0.001 mg/L; 0.01 mg/L
Iron AccuracyLow Range: ±0.01 mg/L ±8% of reading High Range: ±0.04 mg/L ±2% of reading
Iron MethodLow Range: Adaptation of the TPTZ Method High Range: Adaptation of the EPA Phenanthroline method 315B, for natural and treated waters
Magnesium Range0 to 150 mg/L (as Mg 2+ )
Magnesium Resolution1 mg/L
Magnesium Accuracy±5 mg/L ±3% of reading
Magnesium MethodAdaptation of the Calmagite method
Manganese RangeLow Range: 0 to 300 μg/L (as Mn) High Range: 0.0 to 20.0 (as Mn)
Manganese Resolution1 μg/L; 0.1 mg/L
Manganese AccuracyLow Range: ±10 μg/L ±3% of reading High Range: ±0.2 mg/L ±3% of reading
Manganese MethodLow Range: Adaptation of the PAN Method High Range: Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Periodate method
Molybdenum Range0.0 to 40.0 mg/L (as Mo 6+ )
Molybdenum Resolution0.1 mg/L
Molybdenum Accuracy±0.3 mg/L ±5% of reading
Molybdenum MethodAdaptation of the mercaptoacetic acid method
Nickel RangeLow Range: 0.000 to 1.000 mg/L (as Ni) High Range: 0.00 to 7.00 g/L (as Ni)
Nickel Resolution0.001 mg/L; 0.01 g/L
Nickel AccuracyLow range: ±0.010 mg/L ±7% of reading High Range: ±0.07 g/L ±4% of reading
Nickel MethodLow Range: Adaptation of the PAN method High Range: Adaptation of the photometric method
Nitrate Range0.0 to 30.0 mg/L (as NO 3 – – N)
Nitrate Resolution0.1 mg/L
Nitrate Accuracy±0.5 mg/L ±10% of reading
Nitrate MethodAdaptation of the cadmium reduction method
Nitrite RangeFreshwater Low Range: 0 to 600 μg/L (as NO 2 – -N) High Range: 0 to 150 mg/L (as NO 2 – ) Seawater Ultra Low Range: 0 to 200 μg/L (as NO 2 – -N)
Nitrite ResolutionFreshwater: 1 μg/L; 1 mg/L Seawater: 1 μg/L
Nitrite AccuracyFreshwater Low Range: ±20 μg/L ±4% of reading High Range: ±4 mg/L ±4% of reading Seawater ±10 μg/L ±4% of reading
Nitrite MethodAdaptation of the EPA Diazotization method 354.1
Oxygen, Scavenger Range0 to 1000 μg/L (as DEHA) 0.00 to 1.50 mg/L (as Carbohydrazide) 0.00 to 2.50 mg/L (as Hydroquinone) 0.00 to 4.50 mg/L (as ISO-ascorbic acid)
Oxygen, Scavenger Resolution1 μg/L (DEHA); 0.01 mg/L
Oxygen, Scavenger Accuracy±5 μg/L ±5% of reading (DEHA)
Oxygen, Scavenger MethodAdaptation of the iron reduction method
Ozone Range0.00 to 2.00 mg/L (as O 3 )
Ozone Resolution0.01 mg/L
Ozone Accuracy±0.02 mg/L ±3% of reading
Ozone MethodColorimetric DPD Method
Phosphate RangeFreshwater Low Range: 0.00 to 2.50 mg/L (as PO 4 3- ) High range: 0.0 to 30.0 mg/L (as PO 4 3- ) Seawater Ultra Low Range: 0 to 200 μg/L (as P)
Phosphate ResolutionFreshwater: 0.01 mg/L; 0.1 mg/L Seawater: 1 μg/L
Phosphate AccuracyFreshwater Low Range: ±0.04 mg/L ±4% of reading High Range: ±1 mg/L ±4% of reading Seawater Ultra Low Range: ±5 μg/L ±5% of reading
Phosphate MethodFreshwater Low Range: Adaptation of the Ascorbic Acid method Freshwater High Range and Seawater Ultra Low Range: Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Amino Acid method
Potassium Range0.0 to 20.0 mg/L (as K)
Potassium Resolution0.1 mg/L
Potassium Accuracy±3.0 mg/L ±7% of reading
Potassium MethodAdaptation of the Turbidimetric Tetraphenylborate method
Silica RangeLow Range: 0.00 to 2.00 mg/L (as SiO 2 ) High Range: 0 to 200 mg/L (as SiO 2 )
Silica Resolution0.01 mg/L; 1 mg/L
Silica AccuracyLow Range: ±0.03 mg/L ±3% of reading High Range: ±1 mg/L ±5% of reading
Silica MethodLow Range: Adaptation of the ASTM Manual of Water and Environmental Technology, D859, Heteropoly Molybdenum Blue method High Range: Adaptation of the USEPA Method 370.1 and Standard Method 4500-SiO 2
Silver Range0.000 to 1.000 mg/L (as Ag)
Silver Resolution0.001 mg/L
Silver Accuracy±0.020 mg/L ±5% of reading
Silver MethodAdaptation of the PAN method
Sulfate Range0 to 150 mg/L (as SO 4 2- )
Sulfate Resolution1 mg/L
Sulfate Accuracy±5 mg/L ±3% of reading
Sulfate MethodTurbidimetric – Sulfate is precipitated with barium chloride crystals
Zinc Range0.00 to 3.00 mg/L (as Zn)
Zinc Resolution0.01 mg/L
Zinc Accuracy±0.03 mg/L ±3% of reading
Zinc MethodAdaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Zincon method
Input Channels1 pH electrode input and 5 photometer wavelengths
pH Electrodedigital pH electrode (not included)
Logging Typelog on demand with user name and sample ID optional input
Logging Memory1000 readings
ConnectivityUSB-A host for flash drive; micro-USB-B for power and computer connectivity
GLPcalibration data for connected pH electrode
Display128 x 64 pixel LCD with backlight
Battery Type / Life3.7 VDC Li-polymer rechargeable battery / >500 photometric measurements or 50 hours of continuous pH measurement
Power Supply5 VDC USB 2.0 power adapter with USB-A to micro-USB-B cable (included)
Environment0 to 50.0 o C (32 to 122.0 o F); 0 to 95% RH, non-condensing
Dimensions206 x 177 x 97 mm (8.1 x 7.0 x 3.8′′)
Weight1.0 kg (2.2 lbs.)
Photometer/Colorimeter Light Source5 LEDs with 420 nm, 466 nm, 525 nm, 575 nm, and 610 nm narrow band interference filters
Photometer/Colorimeter Light Detectorsilicon photodetector
Bandpass Filter Bandwidth8 nm
Bandpass Filter Wavelength Accuracy±1 nm
Cuvette Typeround, 24.6 mm
Number of Methods128 max.
Ordering InformationHI83300 is supplied with sample cuvettes and caps (4 ea.), cloth for wiping cuvettes, USB to micro USB cable connector, power adapter and instruction manual.