Binocular Microscope Specification (คุณลักษณะเฉพาะ)

A binocular microscope specification for teaching and routine lab work, in 14 lines: 8 mandatory and 6 desired, each with how it is checked at acceptance. Copy it into a TOR or a reference-price file, or download the Excel workbook with the table already filled in. Free, no signup.

Download this specification (.xlsx)

The TOR workbook with this table already filled in: a numbered TOR outline and the specification sheet with level, verification method and a vendor-response column. Bilingual Thai and English; the working sheets carry no branding. Edit freely.

The specification

8 mandatory lines (an offer that fails one is out) and 6 desired lines (scored, not disqualifying).

ItemRequirementLevelVerified by
1. Objectives4x, 10x, 40x and 100x oil-immersion objectives, achromat or better, NA at least 0.10, 0.25, 0.65 and 1.25, on a nosepiece with at least 4 positions. The 100x objective and immersion oil are delivered with every unit.MandatoryDatasheet and the engraving on each objective
2. Optical systemInfinity-corrected optics with plan (flat-field) objectives, and optics treated against fungusDesiredDatasheet
3. EyepiecesOne pair of 10x wide-field eyepieces, field number at least 20, with diopter adjustment on at least one sideMandatoryThe "10x/20" engraving and a demonstration
4. HeadBinocular head inclined between 25 and 45 degrees, with interpupillary adjustment covering at least 55 to 75 mmMandatoryDemonstration with the narrowest- and widest-set eyes among the users
5. StageMechanical stage with a slide holder and position scales, X-Y travel at least 75 x 26 mmMandatoryDatasheet and a demonstration
6. FocusCoarse and fine focus, the fine knob moving at most 0.3 mm per turn or graduated in steps of at most 3 µm, with an upper stop that keeps the objective off the slideMandatoryDemonstration at 40x
7. Classroom handlingScored per feature: coaxial coarse and fine knobs on both sides with tension adjustment, a rackless stage, an inward-facing nosepiece, and secured eyepiecesDesiredDemonstration
8. CondenserAbbe condenser, NA at least 1.25, with an aperture iris diaphragm; height-adjustable or pre-centred at the factoryMandatoryDatasheet and an iris demonstration
9. IlluminationBuilt-in LED with continuous brightness control, running on 220 V 50 Hz, delivered with a cord or adapter that fits Thai socketsMandatoryDemonstration and a count of the cords
10. LED lifeRated LED life at least 20,000 hoursDesiredDatasheet
11. Dust coverOne dust cover per unitDesiredCount
12. WarrantyAt least 24 months on the whole instrument, optics and LED included, from the acceptance dateMandatoryWarranty letter
13. Manual and handoverA Thai-language manual or quick guide, and a hands-on handover to the teaching staff or users on deliveryDesiredCount of manuals and a sign-in sheet
14. After-sales serviceA yearly preventive check during the warranty (cleaning, re-greasing, fungus inspection) and a written commitment to parts and repair for at least 5 yearsDesiredVendor letter

Every line about the instrument is met by current models from at least three makers, checked against the makers’ own documents: Evident CX23, Nikon ECLIPSE Ei, ZEISS Primostar 1, ZEISS Primostar 3, Leica DM500 / DM750

Lines that lock a spec to one brand

Copying one datasheet

Objective working distances, stage dimensions, eyepoint height and a maker’s own feature names each fit one model. Leave them out: they do not matter for teaching or routine work.

Exact ranges instead of "covers at least"

An interpupillary range of exactly 48 to 75 mm excludes microscopes that adjust from 50 to 75 mm and work just as well. Write "covers at least 55 to 75 mm".

Wattage as a stand-in for brightness

An "LED at least 3 W" line excludes 0.5 to 1 W illuminators that are bright enough. Ask for continuous brightness control and check it by demonstration.

Missing deliverables

Some makers sell the 100x oil objective as an option, and immersion oil, dust covers and power cords can be missing too. State that they are delivered. A spare bulb is obsolete on LED stands.

Mixing figures from two datasheets

A spec that takes some lines from one brand and some from another is often met by no microscope at all. Check every line against at least three makers before publishing.

Checks on delivery day

  • Count and read: objectives engraved 4x/0.10, 10x/0.25, 40x/0.65 and 100x/1.25 Oil, eyepieces 10x/20; model and serial numbers match the quotation and the warranty card; oil, cover and cord present.
  • Turn the brightness from minimum to maximum: the change should be smooth, without flicker, and the field evenly lit. Open and close the condenser iris at 40x and watch the contrast change.
  • Focus a stained slide at 10x and switch to 40x: only the fine knob should be needed. Rack the stage up and confirm the stop keeps the objective off the slide, and that the stage does not drift down on its own.
  • Have the users with the narrowest- and widest-set eyes look through it: one merged circle without strain, and a working diopter adjustment.
  • Apply oil and check a sharp image at 100x. Run the stage through its full X and Y travel, then shine a torch into the objectives and eyepieces: no haze or spots, the first signs of fungus.

Record the result on a goods receipt: free goods receipt template

Quotations for the reference price

Send this table, edited to your needs, and we quote against it line by line: brand and model, how each line is met, and the maker’s documents that show it. Any brand on one quotation. Request quotations for a reference price

Models we quote for this instrument: Olympus CX23 · ZEISS Primo Star · Motic SFC-282 · Optec B203 LED

Common questions

Should the spec include objective working distances?

No. They vary by maker and hardly matter for teaching or routine work, so including them matches the spec to one model. The objectives’ NA says more about image quality.

Does a teaching lab need plan objectives?

Not as a requirement. Standard achromats teach well; plan objectives stay sharp to the edge of the field, which helps for whole-field viewing and photography, so the table lists them as desired.

What is the difference between a binocular and a trinocular microscope?

A trinocular head adds a third port for a camera. If you need to project or record images, specify a trinocular head and add the camera as a separate item. The rest of this table still applies.

Why state that the 100x objective is delivered?

Some makers sell the 100x oil objective as an option outside the standard set. If the spec does not require it, some offers include it and others do not, and the prices stop being comparable.