We use cookies to improve this site and see how it is used. Cookies Policy

Skip to content
ไทย

Specification for a TOR or reference price

Laboratory Autoclave Specification (คุณลักษณะเฉพาะ)

A vertical laboratory autoclave specification for media, glassware and waste, 50 L and up, in 14 lines: 8 mandatory and 6 desired, each with how it is checked at acceptance. Copy it into a TOR or a reference-price file, or download the Excel workbook with the table already filled in. Free, no signup.

Download this specification (.xlsx)

The TOR workbook with this table already filled in: a numbered TOR outline and the specification sheet with level, verification method and a vendor-response column. Bilingual Thai and English; the working sheets carry no branding. Edit freely.

The specification

8 mandatory lines (an offer that fails one is out) and 6 desired lines (scored, not disqualifying).

ItemRequirementLevelVerified by
1. Chamber and capacityVertical, top-loading; chamber volume at least 50 L, with usable space at least 300 mm in diameter and 600 mm deepMandatoryDatasheet, and a tape measure on delivery
2. Chamber materialStainless steel chamber, SUS 304 or betterMandatoryDatasheet
3. Temperature and pressureSterilizing temperature adjustable over at least 115 to 132 °C; maximum working pressure at least 0.20 MPaMandatoryDatasheet and a demonstration of the setting
4. ProgramsSeparate solid and liquid programs, the liquid program ending with a slow exhaust or natural cooling so nothing boils over; usable for wasteMandatoryDemonstration
5. Warming and agar dissolvingKeep-warm after sterilizing covering 50 to 60 °C, and an agar-dissolving mode at 100 °C or belowDesiredDemonstration
6. Load temperature probeA flexible probe placed in a sample bottle, with the hold time counted from the load’s temperature (make it mandatory if 1 L media bottles are routine)DesiredDemonstration and the cycle record
7. Cooling and dryingFan-assisted cooling; drying only if glassware must come out dryDesiredDatasheet
8. SafetyLid locked while there is pressure or the temperature is above the release setpoint; safety valve, over-temperature and over-pressure protection, low-water cut-off, earth-leakage breaker and alarm displayMandatoryDemonstration and datasheet
9. Exhaust and drainExhaust steam condensed into a closed bottle or a cooled drain, drain bottle supplied, and the feed water statedDesiredInspection
10. Display and recordsDigital display, pressure gauge, and a cycle record (printer, USB or RS-232) with date, program, temperature and pressureDesiredA record printed during the demonstration
11. Delivered withAt least 2 stainless baskets sized to the chamber, a bottom plate, and manuals in Thai and EnglishMandatoryCount on delivery
12. Power220 to 230 V, 50 Hz, single phase, with the rated current statedMandatoryDatasheet and a site check
13. Warranty and serviceAt least 1 year of warranty; installation, testing and training in Thai by a technician trained by the makerMandatoryWarranty letter and the technician’s training record
14. Documents and commissioningPressure-vessel conformity papers to the maker’s standard, one commissioning cycle with chemical and biological indicators, and calibration certificates for the temperature sensor and pressure gaugeDesiredIndicator results and certificates

Every line about the instrument is met by current models from at least three makers, checked against the makers’ own documents: TOMY SX-300/500/700, Hirayama HV-II, Raypa AES, ALP CL series, Systec VX series

Lines that lock a spec to one brand

A CSSD sterilizer spec for lab work

Pre/post-vacuum sterilizers for wrapped surgical sets need a vacuum pump, plumbed water and Bowie-Dick testing, and handle liquids poorly. Media, glassware and waste need a vertical gravity-displacement autoclave.

Litres alone

Some makers count the whole chamber, others only the usable space. State the volume and the usable diameter and depth together.

Copying one brochure

Exact chamber dimensions, 0.255 MPa, a foot pedal or a heater of at least 2,600 W each fit one model. Use minimums several makers meet.

No liquid program

Without a liquid program or load probe, hold time counts from the chamber temperature: large media bottles are under-processed, and a fast exhaust boils them over.

Ignoring the room’s power

Many makers’ larger autoclaves run on 380 to 400 V three phase. Check the room’s supply and run a dedicated circuit; never a household outlet.

Checks on delivery day

  • Check model, serial number, rating plate and pressure-vessel papers against the quotation; count the baskets and measure the usable space.
  • Interlock test: during a cycle, and after stopping one partway, the lid stays locked until pressure is zero and the temperature is below the release setpoint.
  • Run one liquid cycle with the probe in a 1 L water bottle: the hold starts at load temperature, nothing boils over, and the lid unlocks at the set temperature.
  • Run one solid or waste cycle with a chemical indicator and a biological indicator at the centre of a bag, the slowest point to heat, and accept once the incubated biological indicator passes.
  • Check the cycle record for date, time, program, temperature and pressure; compare it with an independent calibrated thermometer, and test the low-water alarm as the manual describes.

Record the result on a goods receipt: free goods receipt template

Quotations for the reference price

Send this table, edited to your needs, and we quote against it line by line: brand and model, how each line is met, and the maker’s documents that show it. Any brand on one quotation. Request quotations for a reference price

Models we quote for this instrument: TOMY SX-500·See the catalog shelf

Common questions

How does a lab autoclave differ from a hospital CSSD sterilizer?

A lab autoclave displaces air with steam and suits media, glassware and waste in open bags, with a liquid program. A CSSD sterilizer pulls a vacuum to remove air from wrapped surgical sets and textiles, and is not designed for liquids.

What capacity should we specify?

Start from the largest bottle or bag you sterilize per cycle and compare it with the usable dimensions, not litres alone. For the 75 L class, ask for at least 75 L and a usable diameter of at least 360 mm.

Do we need drying?

Not for media and waste. Add it only if glassware or instruments must come out dry and ready to use at the end of the cycle.

Why test with a biological indicator on delivery?

A biological indicator shows the cycle actually sterilizes at the slowest point to heat; a chemical indicator shows the set conditions were reached. The delivery test is the baseline for later periodic checks.

Other ready-written specifications: Binocular microscope · Analytical balance (0.1 mg)·The blank TOR workbook

Specification ready? Get quotations against it

Paste or upload the table and we answer it line by line, with the maker documents your acceptance check will ask for.